Technical data

42 Description
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Halogen illumination system
The stand is equipped with two separate illumination systems for fiber il-
lumination. Each of the two lamp housings contains a backup lamp which
will be automatically activated if the first lamp fails. Only one of the two
illumination systems is required for normal operation of a surgical micro-
scope. The second illumination system can be used, for example, for a
fiber slit lamp or a dual fiber illumination system.
1 Lamp module
2 Ventilation grid
Do not cover the ventilation grid! Make sure that drapes do not cover
the grid. This can lead to overheating of the lamp modules and to lamp
failure.
3 Flap
The flap is the mechanical indicator for the operating status of the hal-
ogen lamps.
When the flap is closed, the main lamp is operative (green light (9)
is on).
When the flap is open, the main lamp has failed. The backup lamp
is operative (yellow light (8) is on).
4 Manual activation of the backup lamp
If the automatic activation system fails, press this button to switch on
the backup lamp.
5 Opening the lamp module
When you press this button, the lamp module is slightly ejected. Pull
out the lamp module all the way for lamp change.
6 Filter knobs
The filter knobs have four positions:
0 no filter
1 GG 475 filter: to protect the patient's eye during surgery against
unnecessary (blue) radiation (retinal injury).
2 KK 40 filter: to increase color temperature
3 no filter