Technical data

54 Preparations for use
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Screw in the securing screw (7) and tighten it firmly . The securing
screw (7) must go into the groove (9). This is ensured when the secu-
ring screw is flush with the outer surface of the coupling.
Hook the coupling (12) including the surgical microscope from above
into the receptacle (11) on the suspension arm and tilt the coupling
downward into its vertical position.
Firmly tighten the mounting screw (3) using a 5 mm Allen key.
Plug the cable clip (10) into the opening (13) of the coupling.
Then perform the balance setting procedure.
Warning!
Before using and after re-equipping the unit, always make sure that se-
curing screws (3), (4) and (7) have been tightened firmly.
Note:
When mounting surgical microscopes with an integrated coupling, steps
3 to 9 need not be performed.