Technical data

60 Preparations for use
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Connecting the stand
Check the voltage indicated at (2).
Caution:
The voltage of the stand is set at the factory to the voltage used in the
country of destination. The rated voltage indicated at window (2) must cor-
respond to the rated voltage available at the site of installation. If this is
not the case, re-adjust the sliding switch using a suitable tool.
Note:
Make sure that power switch (4) is off before inserting connectors in or re-
moving them from (1) und (3).
Plug the connector of the foot control panel or operating chair into con-
nector (1) of the stand.
Install the microscope cable in the cable clip in such a way that the ca-
ble is neither stretched nor kinked when the microscope is turned or
tilted.
Install the light guide in the cable clip and insert its end into the recep-
tacle on the microscope as far as it will go. Make sure that the light
guide is not stretched or kinked when the microscope is turned or tilt-
ed.
Connect the stand to line power using the power cord supplied for this
purpose. Only use power outlets which are provided with a properly con-
nected protective earth connector.