Technical data

66 Operation
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Adjusting the limit of downward movement
The vertical lift of the suspension arm must be limited in such a way that
the patient's safety is also ensured when the microscope is inadvertently
lowered.
Loosen the adjustment screw (1) by a few turns.
Press one of the release keys for the magnetic clutches on the surgi-
cal microscope and lower the microscope to a position where it can be
focused on the surgical field (depending on the focal length of the ob-
jective lens) and where the safety distance to the surgical field is still
sufficient.
Turn the adjustment screw (1) clockwise as far as it will go.
Move the surgical microscope again to the lower limit and check the
safety distance.