Acoustic signals
Service display
In the event of an error, e.g. during the software check following the
startup of the ceiling mount (floor stand), the display and key field displays
an error message in the form of a wrench symbol, accompanied by three
successive beeps.
Motorized control of the surgical microscope is not possible in this case.
With the exception of the recentering of the X-Y coupling, all
other functions of the surgical microscope can only be manually operated.
You can still release the magnetic clutches using the appropriate key in
the left or right handgrip of the surgical microscope.
If you press the "MANUAL" key, the surgical microscope can no longer be
operated via the foot control panel, the handgrips or the display and key
field.
You can continue to use the illumination system. The lamp brightness,
however, is automatically set to a fixed value. You can switch the illumi-
nation on and off on the foot control panel.
Three succes-
sive beeps
Error message during software check after switch-
ing on the ceiling mount (floor stand).
Error message for an internal system error.
Single beep When the focus and zoom positions are stored.
Single beep Only with OPMI
®
VISU on the ceiling mount (floor
stand):
when brightness level 1.0 has been reached.
Single beep Only with OPMI
®
VISU on the ceiling mount (floor
stand):
after switching on the ceiling mount (floor stand), if the
lamp brightness of one illumination system has been
set to level 1.0 or higher.