Technical data

84 Operation
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
OPMI
®
Vario: assigning a function to the freely configurable keys on
the handgrips
In configuration mode 1, you can use the bottom section of the display
and key field (HAND GRIP:) to assign any of the following functions to the
keys in the handgrip:
Rollover
Select the function required using a rollover run in the sequence defined
in the above table.
At each press of the "" key, you advance in clockwise direction. At each
press of the "" key, you advance in counterclockwise direction:
XY-RES LIGHT
↑↓
PH/AUX ¨ MEM
If you have assigned the memory function (MEM) to keys (1) and (2) of
the handgrips, you can determine by the length of the key pressure during
operation whether a positioning run is to be triggered or whether the cur-
rent position is to be saved.
If you press key (1) for less than 2 seconds, a focus positioning run is
started; if you press this key longer than 2 seconds, the current focus po-
The two upper keys (1 and 2) in the left and right handgrips of the OPMI
®
Vario surgical microscope are freely configurable.
1
2
3
Display Function of left key (1) Function of right key (2)
LIGHT Reduce brightness Increase brightness
MEM Focus memory
Key pressed < 2 sec:
triggers positioning
Key pressed > 2 sec:
saves the position
Zoom memory
Key pressed < 2 sec:
triggers positioning
Key pressed > 2 sec:
saves the position
PH/AUX Triggers an exposure at the
photo interface
Triggers an AUX signal at the
AUX interface
XY-RES XY recentering ("Reset") XY recentering ("Reset")
Path: After you have switched on the instrument, the basic mode is auto-
matically displayed.
To access configuration mode 1 from the basic mode, simultane-
ously press the "MODE" and "STORE" keys.
MODE
STORE MANUAL
FOOT
LEFT C:
PHOTO
FOOT
RIGHT D:
PHOTO
HAND
GRIP:
LIGHT
Configuration
mode 1