Technical data
84 Operation
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
OPMI
®
Vario: assigning a function to the freely configurable keys on
the handgrips
In configuration mode 1, you can use the bottom section of the display
and key field (HAND GRIP:) to assign any of the following functions to the
keys in the handgrip:
Rollover
Select the function required using a rollover run in the sequence defined
in the above table.
At each press of the "∇" key, you advance in clockwise direction. At each
press of the "∆" key, you advance in counterclockwise direction:
→ XY-RES → LIGHT →
↑↓
← PH/AUX ¨ MEM ←
If you have assigned the memory function (MEM) to keys (1) and (2) of
the handgrips, you can determine by the length of the key pressure during
operation whether a positioning run is to be triggered or whether the cur-
rent position is to be saved.
If you press key (1) for less than 2 seconds, a focus positioning run is
started; if you press this key longer than 2 seconds, the current focus po-
The two upper keys (1 and 2) in the left and right handgrips of the OPMI
®
Vario surgical microscope are freely configurable.
1
2
3
Display Function of left key (1) Function of right key (2)
LIGHT Reduce brightness Increase brightness
MEM Focus memory
Key pressed < 2 sec:
triggers positioning
Key pressed > 2 sec:
saves the position
Zoom memory
Key pressed < 2 sec:
triggers positioning
Key pressed > 2 sec:
saves the position
PH/AUX Triggers an exposure at the
photo interface
Triggers an AUX signal at the
AUX interface
XY-RES XY recentering ("Reset") XY recentering ("Reset")
Path: After you have switched on the instrument, the basic mode is auto-
matically displayed.
To access configuration mode 1 from the basic mode, simultane-
ously press the "MODE" and "STORE" keys.
MODE
STORE MANUAL
FOOT
LEFT C:
PHOTO
FOOT
RIGHT D:
PHOTO
HAND
GRIP:
LIGHT
Configuration
mode 1