Technical data

90 Operation
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
OPMI
®
Vario: defining the power-on instrument behavior
The power-on mode permits you to define how the instrument should be-
have immediately after it has been switched on.
Use the bottom section (POW ON MODE:) of the display and key field in
configuration mode 2 to select one of the following power-on modes:
Rollover
Select the function required using a rollover run in the sequence defined
in the above table.
At each press of the "" key, you advance in clockwise direction. At each
press of the "" key, you advance in counterclockwise direction:
MEM OFF
↑↓
PRESET
Caution:
Be extremely careful when changing these settings. You should change
settings only under your own user ID. Notify all users of any changes, or
make sure that each user only works under his own user ID.
OFF No positions are changed on instrument startup.
After switching on the instrument, you can use the appro-
priate keys on the foot control panel or handgrips to set the
focus and zoom manually to the user-specific values stored
in the memory function.
PRESET On instrument startup, the focus (working distance) and
zoom are automatically set to the factory-adjusted values:
focus = 250 mm and zoom = 0.5.
These two values are also stored in the memory function
(FOC-MEM and ZOOM-MEM).
MEM On instrument startup, the focus (working distance) and
zoom are automatically set to the user-specific focus and
zoom settings stored in the memory functions. The values
in the FOC-MEM and ZOOM-MEM memory functions were
previously stored for the respective user by pressure of the
"STORE" key.
Path: The basic mode is automatically displayed after the instrument has
been switched on.
To access configuration mode 2 from the basic mode, go via con-
figuration mode 1.
For this, simultaneously press the "MODE" and "STORE" keys
twice in succession:
1. Jump from the basic mode to configuration mode 1
2. Jump from configuration mode 1 to configuration mode 2.
MODE
STORE MANUAL
SPEED
FUNCT:
MEDIUM
LIGHT
FUNCT:
MANUAL
POW ON
MODE:
MEM
Configuration
mode 2