Technical data
92 Operation
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
OPMI
®
Vario: entering the eyepiece magnification for the computati-
on of the total magnification
In the basic mode, the total magnification (MAG) of the optical system can
be displayed as a factor in the middle section of the display and key field.
The total magnification is computed by the software. The magnification
data of the eyepieces currently used must be known to the software for
this purpose.
Use the top section (EYEPIECE:)of the display and key field in configura-
tion mode 3 to select one of the possible eyepiece magnifications:
Rollover
Select the function required using a rollover run in the sequence defined
in the above table.
At each press of the "∇" key, you advance in clockwise direction. At each
press of the "∆" key, you advance in counterclockwise direction:
→ 16x → 10x →
↑↓
← 12,5x ←
Saving parameter settings
As soon as you have entered a parameter setting, it is saved under the
current user ID.
Caution:
Be extremely careful when changing these settings. You should change
settings only under your own user ID. Notify all users of any changes, or
make sure that each user only works under his own user ID.
10x
12.5x
16x
Path: The basic mode is automatically displayed after the instrument has
been switched on.
To access configuration mode 3 from the basic mode, go via con-
figuration modes 1 and 2.
For this, simultaneously press the "MODE" and "STORE" keys
three times in succession:
1. Jump from the basic mode to configuration mode 1,
2. Jump from configuration mode 1 to configuration mode 2,
3. Jump from configuration mode 2 to configuration mode 3.
MODE
STORE MANUAL
EYE-
PIECE:
10x
SELECT
BRAKES:
ALL
Configuration
mode 3