Operation Manual

6000 COUNTS WITH APP
DC/AC TRUE-RMS CLAMP MULTIMETER
OPERATION MANUAL
1. Overview
The auto range clamp multimeter is a portable and stable performance. Using
6000 counts digit LCD monitor with character 16mm high. With overall circuitry
design centering on large-scale IC A/D converters in conjunction and over-load
protection circuit, the meters give excellent performance and exquisite making as
a handy utility instrument.
The meter can be connected with mobile phone by Bluetooth, and display on
phone by APP, you can remote monitoring the measurement condition, the
distance control is 1015m.
The meters can be used to measure DC & AC voltage, DC & AC current,
resistance, capacitor, frequency, duty cycle, temperature, Non Contact AC
Voltage (NCV) detection, positive diode voltage fall and audible continuity.
2. Panel Layout
1. Clamp jaws: Opens 26mm to enclose conductor.
2. NCV detection area (bottom).
3. Lamp light: Press the HOLDkey over 2 seconds to light the lamp light, it will
light up too when the built-in buzzer sounds.
4. Jaw-opening handle: Opens and closes the jaws.
5. Rotary Switch: Use this switch to select functions and ranges.
6. HOLD key: Press theHOLD” key to lock display value, and the “DH sign will
appear on the display, press it again to exit. PressHOLD” key more than 2
seconds, the back light and the lamp light will light up, press it more than 2
se
conds again, the back light and the lamp light will light off.
7. SEL key: This key work on the "
" range, press the key to choose
resistance, continuity, diode or capacitance test, on the voltage or current range,
change to DC or AC, on the range, change to or test .
8. MAX/MIN key: Press the MAX/MINkey to lock MAX or MIN value, and the
MAX” or “MIN” sign will appear on the display, press it over 2 seconds to exit.
9. REL Key: Press theREL key, the meter enters relative measuring mode,
is displayed on the LCD and the present reading becomes the reference value
and displayed on the display. Relative measurement REL=measurement
value-Reference value. Press it again to exit.
10. Hz/Duty Key: On ACV/ACAor Hzrange, press the Hz%key, you can
choose the Frequency or Duty Cycle measurement.
11. LCD display: 6000 counts digit, full function symbol display.
12. COM: COM and Temperature “- Input Jack
13. V
Hz : V/ / /Hz/T+ Input Jack
3. Safety Information
3-1 The meter is designed according to IEC-1010 concerning electronic
measuring instruments with an over-voltage category 600V (CAT ) and
pollution 2.
3-2 Follow all safety and operating instructions to ensure that the meter is used
safely and is kept in good operating condition.
3-3 safety symbols:
Important safety information, refer to the operating manual.
Dangerous voltage may be presence.
Double insulation (protection Class II)
4. Special Cautions for Operation
4-1 The meters can be safe only according to standard procedures when used in
conjunctions with the supplied test leads. To replace damaged test leads with only
the same model or same electric specifications.
4-2 To avid risk of electric shock, do not use the meters before the cover is in
place.
4-3 The range switch should be right position for the testing.
4-4 To avoid electric shock and damaging the instruments, the input signals are
forbidden to exceed the specified limits.
4-5 When measuring TV set or switched power, attention should be paid to the
possible pulses that may bring destruction to the circuit.
4-6 Range switch position is forbidden to be changed at random during
measurement.
4-7 Take caution against shock in the course of measuring voltage higher than
DC 60V & AC 30V.
4-8 Before opening the cover of the battery cabinet to replace batteries.
disconnect the test leads from any external circuit, set the selector switch to
"OFF" position.
4-9 Keep the fingers after the protection ring when measuring through the
instrument lead.
4-10 Keep the fingers after the protection ring when measuring through the
clamp.
4-11 After operation is finished, set function switch at OFF to save battery power.
4-12 If the meter is without usage for long time, take out battery to avoid damage
by battery leakage.
5. GENERAL SPECIFICATIONS
5-1 Max Voltage between input terminal and Earth Ground:
CAT 600V or CAT 300V
5-2 Over-range Indication: display “OL” for the significant digit.
5-3 Automatic display of negative polarity “-” .
5-4 Low Battery Indication:
displayed.
5-5 Max LCD display: 6000 counts digit.
5-6 Auto range control
5-7 Clamp opening size: 26mm.
5-8 Power supply: 1.5V×2 “AAA” R03P battery
5-9 Operating Temp.: 0 to 40 (relative humidity <85%)
5-10 Storage Temp.:-10 to 50 (relative humidity <85%)
5-11 Guaranteed precision Temp.: 23±5 (relative humidity <70%)
5-12 Dimension: 207(H)×75(W)×37(D)mm.
5-13 Weight: Approx. 280g (including battery).
6. Testing Specifications
Accuracy is specified for a period of year after calibration and at 18 to 28
(64 to 82 ) with relative humidity to 70%.
6-1 DC Voltage
Range
Resolution Accuracy
600mV
0.1mV
6V 1mV
60V 10mV
±(0.5% of rdg + 2 digits)
600V 100mV ±(0.8% of rdg + 2 digits)
-- Impedance: 10M, More than 100M on 600mV range
-- Overload protection: 600V DC or AC rms
6-2 AC Voltage (True RMS)
Range
Resolution Accuracy
6V 1mV
60V 10mV
±(1.0% of rdg + 3 digits)
600V 100mV ±(1.5% of rdg + 3 digits)
-- Impedance: 10M
-- Overload protection: 600V DC or AC rms
-- Frequency Range: 40 to 2kHz
6-3 DC Current
Range
Resolution Accuracy
60A 10mA ±(2.5% of rdg + 10 digits)
400A 100mA ±(3.0% of rdg + 10 digits)
-- Overload protection: 400A DC or AC rms
6-4 AC Current (True RMS)
Range
Resolution Accuracy
60A 10mA ±(2.5% of rdg + 10 digits)
400A 100mA ±(3.0% of rdg + 10 digits)
-- Overload protection: 400A DC or AC rms
-- Frequency Range: 40 to 100Hz
6-5 Resistance
Range
Resolution Accuracy
600 0.1 ±(1.0% of rdg + 3 digits)
6k 1
60k 10
600k
100
6M 1k
±(1.0% of rdg + 2 digits)
60M 10k ±(1.5% of rdg + 3 digits)
-- Overload protection: 250V DC or AC rms
6-6 Capacitance
Range Accuracy Resolution
6nF ±(5.0% of rdg + 10 digits) 1pF
60nF 10pF
Fp001 Fn006
6µF
±(3.0% of rdg + 10 digits)
1nF
60µF ±(5.0% of rdg + 10 digits) 10nF
600µF 100nF
Fµ1 Fm6
60mF
±(10.0% of rdg + 20 digits)
10µF
-- Overload protection: 250V DC or AC rms
6-7 Frequency
Range Accuracy Resolution
9.999Hz 0.001Hz
99.99Hz 0.01Hz
999.9Hz 0.1Hz
zH1 zHk999.9
99.99kHz 10Hz
999.9kHz 100Hz
9.999MHz
± (0.1% of rdg + 5 digits)
1kHz
-- Sensitivity: sine wave 0.6V rms (9.999MHz: 1.5V rms)
-- Overload protection: 250V DC or AC rms
6-8 Duty cycle
0.1%99.9%: ± ( 2.0% of rdg + 2 digits ), Frequency lower than 10kHz
-- Sensitivity: sine wave 0.6V rms
-- Overload protection: 250V DC or AC rms