R Series® Configuration Manual 9650-0902-01 Rev.
The issue date for the R Series Configuration Guide (REF 9650-0902-01 Rev. K) is May, 2012. If more than 3 years have elapsed since the issue date, contact ZOLL Medical Corporation to determine if additional product information updates are available. © 2012 ZOLL Medical Corporation. All rights reserved. R Series, pedi-padz, Real CPR Help, Code-Ready, and OneStep are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and/or other countries.
Table of Contents Chapter 1 Introduction Entering System Configuration Mode .................................................................................... 2 Printing Configuration Settings and Activity Logs .................................................................. 3 Uploading Data from a Compact Flash Card ......................................................................... 6 Saving Configuration to a Compact Flash Card...................................................................
TABLE OF CONTENTS Advisory Defib...................................................................................................................... 24 Energy Level: Shock 1 ................................................................................................. 24 Energy Level: Shock 2 ................................................................................................. 25 Energy Level: Shock 3 ..........................................................................................
NIBP Settings (Optional)...................................................................................................... 43 Blood Pressure Units ................................................................................................... 43 Trigger NIBP Meas on BP Alarm ................................................................................. 43 Trigger NIBP Meas on HR Alarm ................................................................................. 43 NIBP Auto Interval Default .
TABLE OF CONTENTS Wi-Fi Settings (Optional)...................................................................................................... 53 Full Disclosure Transfer Mode ..................................................................................... 53 Defib History Transfer Mode ........................................................................................ 53 Infrastructure Mode Options ........................................................................................
Chapter 1 Introduction The R Series® is designed with several user selectable features that allow operators to configure the device according to their protocols and local requirements. This manual describes the configuration settings (options and defaults) and how to change them for ALS, Plus, and BLS models. References to AED settings are for Plus and BLS models only. Note: Your unit may not contain all features listed. For BLS/Plus units, some features are available in manual mode operation only.
Chapter 1 Introduction Entering System Configuration Mode In order to make any changes to unit configuration, you must first power on the unit in System Configuration mode. Procedure Follow these steps to power on the unit in System Configuration mode: Step Action 1. Turn the selector switch to OFF, and wait 10 seconds.
Printing Configuration Settings and Activity Logs Printing Configuration Settings and Activity Logs To print the current configuration or activity log of your unit while the unit is in the System Configuration mode, press the Print softkey from the Main System Configuration menu. For instructions on putting the unit in System Configuration mode, refer to the previous section, “Entering System Configuration Mode.
Chapter 1 Introduction 4 9650-0902-01 Rev.
Printing Configuration Settings and Activity Logs Printing the Activity Log In System Configuration mode, press the Print softkey, then the Print Activity Log softkey to print a stripchart of the unit’s activity log. The log consists of the last 1,024 user entries or system prompts. See a copy of a sample Activity Log report below. 9650-0902-01 Rev.
Chapter 1 Introduction Uploading Data from a Compact Flash Card You can upload data from a compact flash card to perform the following tasks: • Copy the configuration of one unit to any number of units • Install new software to the unit • Copy a second language to the unit Procedure Follow these steps to load a configuration, another language, or software from a compact flash card: Step Action 1.
Uploading Data from a Compact Flash Card System Configuration mode, press the Exit Config softkey. See “Leaving Configuration Mode” on page 11 for more information. If the unit encounters an error reading the memory card, one of the following error messages is displayed: Error Message Description CARD READ FAILED No response is detected from the card reader. BAD CARD The unit cannot read the card. NO CARD INSERTED The card is not installed in the data card slot. 9650-0902-01 Rev.
Chapter 1 Introduction Saving Configuration to a Compact Flash Card Whenever you exit configuration mode, you can optionally save the configuration to a compact flash card. This allows you to copy the configuration from one unit to any number of units. Procedure Follow these steps to save configuration data to a compact flash card: Step Action 1. Power on the unit in System Configuration mode as described in “Entering System Configuration Mode” on page 2. System Configuration Print 2.
Restoring Default Values Restoring Default Values You can restore all configuration settings to their factory default settings. Only the display language, if different from the default, does not change. Procedure Follow these steps to restore all configuration settings to default values: Step Action 1. Power on the unit in System Configuration mode as described in “Entering System Configuration Mode” on page 2. System Configuration Print 2.
Chapter 1 Introduction Accessing and Changing Configuration Settings You can change the current value for any configuration setting. You access most of the settings in the same way as described below. Additional steps are required to access and modify several settings. For those settings, the additional steps are provided with the setting description in the next chapter. Procedure Follow these steps to access the settings. Step Action 1.
Leaving Configuration Mode Step Action 4. When you are finished, you can either press the Category List softkey to return to the Category screen, or press the Prev Item or Next Item softkey to move to another option whose settings you want to modify. Leaving Configuration Mode Procedure Follow the step below to leave System Configuration mode: Step Action 1. Press the Exit Config softkey from the Main System Configuration screen to display the Exit Configuration screen.
Chapter 1 Introduction 12 9650-0902-01 Rev.
Chapter 2 Configuration Settings This chapter describes all of the configuration settings on the R Series and includes the following sections.
Chapter 2 Configuration Settings General Settings General settings are attributes that affect the unit’s basic operation and accessibility and include: • Device Identifier • Configuration Mode Passcode • Code Markers Device Identifier The Device Identifier feature allows you to set an 11-digit device identifier code. This code is divided into two sub codes: the site ID (consisting of eight alphanumeric characters) and the device ID (consisting of three alphanumeric characters).
General Settings Step Action 3. In the character selection table: • lets you move to the next or previous characters. Use the softkeys as follows: — moves the highlight up, down, left, and right through the character selection table. • Enter Char — enters the highlighted character for 8 digits of the Site ID and the 3 digits of the Device ID. • Return — returns you to the General Settings screen. Press the Return softkey when you are finished entering the IDs.
Chapter 2 Configuration Settings Configuration Mode Passcode This feature allows you to set an 8-digit system configuration passcode. Once set, only those people who know the passcode can access the configuration mode and modify settings. Values are as follows: Possible Values: 0 through 3 inclusive for each digit Default: 00000000 Procedure Follow these steps to set up a configuration mode passcode: Step Action 1.
General Settings Code Markers This setting allows you to review and change the preconfigured list of Code Markers for use in DEFIB, MONITOR, and PACE modes. Code Marker settings are language dependent. You can specify up to three groups (MONITOR, DEFIB, and PACE) of 20 entries each. Procedure Follow these steps to review, change, or reset code marker settings: Step Action 1. From the Select Configuration Categories screen, select General, then press the Enter softkey.
Chapter 2 Configuration Settings Step Action 4. Use the softkeys as described to modify code marker items as needed: • Prev Item and Next Item — scrolls the highlight to the next or previous code marker item. • Change Value — allows you to pick a new marker for the selected item number. When you press the Change Value softkey, three new softkeys become available that allow you to highlight and select the code marker treatment for this item.
ECG Settings ECG Settings ECG monitoring settings are attributes associated with using the R Series unit for ECG monitoring, and include: • Notch Filter • Display Lead on Power-Up • Frequency Response • Gain on Power Up • Enable Leads Off • Enable Pacer Detection Notch Filter This setting allows you to select the ECG Notch Filter frequency for proper ac mains interference rejection.
Chapter 2 Configuration Settings Gain on Power Up This setting allows you to select the default size of the displayed ECG signal when you power on the unit. Values are as follows: Possible Values: 0.5 cm/mV, 1.0 cm/mV, 1.5 cm/mV, and 2.0 cm/mV Default: 1.
Audio Settings Audio Settings The Tone settings control audio functions on the R Series unit, and include: • QRS Volume on Power-Up • Low Battery Tone • Enable Corrective Voice Prompts QRS Volume on Power-Up This setting allows you to select the default QRS beeper volume when you power on the unit. If you set this setting to Off, the unit does not beep during ECG monitoring.
Chapter 2 Configuration Settings Basic Defibrillation Settings The Basic Defibrillation settings control functions related to performing defibrillation with the R Series unit in Manual mode and include: • Defib Default to Pads • Basic Auto Energy Escalation • Retain Sync After Defib • Defib Ready Hold Time • Enable Remote Sync Defib Default to Pads This setting allows you to override the “Display Lead on Power-Up” setting and use Pads/ Paddles whenever the unit is switched to DEFIB mode.
Basic Defibrillation Settings Defib Ready Hold Time This setting allows you to specify the amount of time that the R Series unit will hold a charge before disarming. Values are as follows: Possible Values: 60 secs, 120 secs Default: 60 secs Enable Remote Sync This setting allows you to enable and disable the Remote Sync feature. If disabled (No), you can perform synchronized cardioversion using only the R Series ECG signal.
Chapter 2 Configuration Settings Advisory Defib The Advisory Defibrillation settings control functions related to performing defibrillation with an R Series unit in Advisory mode and include: • Energy Level: Shock 1 • Energy Level: Shock 2 • Energy Level: Shock 3 • Ped Energy Level Shock 1 • Ped Energy Level Shock 2 • Ped Energy Level Shock 3 • Auto Analyze • Enable Check Pulse Prompt • Check Patient Prompt • Auto Charge in Advisory Mode • Enable Voice in Advisory Mode • Display DO CPR Message • CPR Messag
Advisory Defib Energy Level: Shock 2 This setting allows you to select the energy level used for the second shock delivered by the defibrillator when the unit is: • in AED mode or Advisory mode and using pads • in Manual mode, using external paddles or pads, and Basic Auto Energy Escalation is set to Yes Values are as follows: Possible Values: Adult: 1-10, 15, 20, 30, 50, 75, 100, 120, 150, and 200 J Default: Adult: 150 J Energy Level: Shock 3 This setting allows you to select the energy level used fo
Chapter 2 Configuration Settings Ped Energy Level Shock 2 This setting allows you to select the energy level used for the second pediatric shock delivered by the defibrillator when the unit is: • in Advisory mode and using OneStep Pediatric electrodes.
Advisory Defib Enable Check Pulse Prompt This setting works in conjunction with the Auto Analyze function and determines whether the unit displays the prompt CHECK PULSE for 10 seconds immediately after delivery of the final shock in a stacked shock sequence, after each No Shock Advised analysis result, or both. After CHECK PULSE clears, the CPR interval begins. When set to No, the unit immediately begins the CPR interval following delivery of the final shock or after each No Shock Advised analysis result.
Chapter 2 Configuration Settings Enable Voice in Advisory Mode This setting allows you to enable or disable voice prompts when using shock advisory functions (ECG analysis) in Manual mode. If enabled, the voice prompts sound at the appropriate time. If disabled, the unit emits two beeps instead of the voice prompts in Manual mode. Voice prompts are in the language specified by the Language option.
Advisory Defib — If the defibrillator disarms itself after it was charged (because the Defib Ready time elapsed before a shock was delivered), the unit displays the PRESS ANALYZE prompt until the ANALYZE softkey is pressed. Note: The Display DO CPR Message feature is operational only when the Auto Charge in Advisory Mode and Auto Analyze options are both set to Yes.
Chapter 2 Configuration Settings Stacked Shocks The setting specifies the number of stacked shocks that must be administered prior to the activation of a CPR interval and works in conjunction with the Auto Analyze setting. If set to 3, the unit will auto analyze 3 times before prompting the user. If set to 1, the unit will auto analyze once.
CPR Settings CPR Settings The CPR settings control functions related to Real CPR Help® function: • Enable CPR Voice Prompts • Enable CPR Display • Enable CPR Metronome • Target CPR Depth • Display Fully Release Prompt Enable CPR Voice Prompts This setting allows you to enable (Yes) or disable (No) voice prompts during CPR. Voice prompts are in the language specified by the Language option. Note: This setting is always set to Yes in AED mode.
Chapter 2 Configuration Settings parameter is set to 2.0 inches, the reference markers on the CPR bar graph are displayed at 2.0 and 2.4 inches. Values are as follows: Possible Values: 1.5 inches, 2.0 inches Default: 1.5 inches Display Fully Release Prompt This parameter, when set to Yes, specifies that the FULLY RELEASE prompt will be displayed every 45 seconds during CPR. If this parameter is set to No, the prompt is not displayed.
Pace Settings Pace Settings The Pace settings are attributes that control functions related to performing pacing with an R Series unit and include: • Pace Rate Setting on Power-Up • Async Softkey in Pace Mode Pace Rate Setting on Power-Up This setting allows you to set the default pace rate, pulses per minute (ppm), of the R Series when you power on the unit.
Chapter 2 Configuration Settings Alarms Settings Alarm settings are attributes associated with Alarm states and thresholds for the R Series unit and include: • Alarm Limits at Power Up • Alarms Active at Power Up 34 9650-0902-01 Rev.
Alarms Settings Alarm Limits at Power Up This option allows you to enable or disable, and set high and low threshold values for the ECG Heart Rate, and the SpO2 alarms on the R Series unit. All alarms listed below (Heart Rate and optional devices SpO2, EtCO2, NIBP) are enabled by default.
Chapter 2 Configuration Settings Possible Values: NIBP Dia: Adult: High: 50-200 mmHg in increments of 5 6.7-26.7 kPa in increments of 0.67 Low: 25-100 mmHg in increments of 5 3.3-13.3 kPa in increments of 0.67 Pediatric: High: 50-130 mmHg in increments of 5 6.7-17.3 kPa in increments of 0.67 Low: 20-100 mmHg in increments of 5 2.7-13.3 kPa in increments of 0.67 Neonatal: High: 30-105 mmHg in increments of 5 4.0-14.0 kPa in increments of 0.67 Low: 20-100 mmHg in increments of 5 2.7-13.
Alarms Settings Default: Heart Rate: High: 150 BPM Low: 30 BPM Optional Devices: SpO2: High: 100% Low: 85% EtCO2: High: 55 mmHg 7.3 kPa 7.2% Low: 25 mmHg 3.3 kPa 3.2% Resp Rate: High: 120 Low: 5 NIBP Sys: Adult: High: 160 mmHg 21.3 kPa Low: 90 mmHg 12.0 kPa Pediatric: High: 145 mmHg 19.3 kPa Low: 75 mmHg 10.0 kPa Neonatal: High: 100 mmHg 13.3 kPa Low: 50 mmHg 6.7 kPa 9650-0902-01 Rev.
Chapter 2 Configuration Settings Default NIBP Dia: Adult: High: 110 mmHg 14.7 kPa Low: 50 mmHg 6.7 kPa Pediatric: High: 100 mmHg 13.3 kPa Low: 35 mmHg 4.7 kPa Neonatal: High: 70 mmHg 9.3 kPa Low: 30 mmHg 4.0 kPa NIBP Mean: Adult: High: 130 mmHg 17.3 kPa Low: 60 mmHg 8.0 kPa Pediatric: High: 110 mmHg 14.7 kPa Low: 50 mmHg 6.7 kPa Neonatal: High: 80 mmHg 10.7 kPa Low: 35 mmHg 4.7 kPa 38 9650-0902-01 Rev.
Alarms Settings Procedure Follow these steps to change the state or value of the ECG Heart Rate alarms: Step Action 1. From the Select Configuration Categories screen, select Alarms, then press the Enter softkey. System Configuration The Alarms Setting screen appears. Alarm Settings Set/Review Set/Review ALARM LIMITS AT POWER UP No ALARMS ACTIVE AT POWER UP Page 1 of 1 Change Value 2. Select Alarm Limits at Power Up, and press the Change Value softkey.
Chapter 2 Configuration Settings Values are as follows: Possible Values: Yes, No Default: No Display Settings The Display settings are attributes that control what and how things appear on the R Series monitor and include: • Display Time • Select Trace 2 At Power Up • Select Trace 3 At Power Up • ECG Color • SpO2 Color • CPR Color • EtCO2 Color • NIBP Color • SPO2 at Power Up Display Time This setting specifies whether to display the elapsed time since the unit was turned on (Elapsed Time), or the rea
Display Settings Select Trace 3 At Power Up This setting allows you to select which waveform will appear in the Trace 3 position on the display when you turn on the R Series unit. Values are as follows: Possible Values: Off, 3 Lead ECG, CPR, Filtered ECG, SpO2, EtCO2 Default: Off Note: When you select 3 Lead ECG, the Select Trace 2 At Power Up setting automatically defaults to 3 Lead ECG.
Chapter 2 Configuration Settings EtCO2 Color This setting allows you to specify the color of the CO2 indicator and waveform (if installed) on the R Series monitor. Values are as follows: Possible Values: Blue, Green, Cyan, Red, Purple, Yellow, White Default: Yellow NIBP Color This setting allows you to specify the color of the NIBP indicator (if installed) on the R Series monitor.
NIBP Settings (Optional) NIBP Settings (Optional) NIBP configuration parameters are available for update only if the NIBP monitoring option is installed on the unit, and include: • Blood Pressure Units • Trigger NIBP Meas on BP Alarm • Trigger NIBP Meas on HR Alarm • NIBP Auto Interval Default • Adult Initial Cuff Inflation • Pediatric Initial Cuff Inflation • Neonatal Initial Cuff Inflation • NIBP Patient Mode at Power Up • NIBP Button Press & Hold • Enable Motion Tolerance Blood Pressure Units This para
Chapter 2 Configuration Settings NIBP Auto Interval Default This parameter allows you to set the default for the amount of time between automatic measurements. Values are as follows: Possible Values: 2.5, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 45, 60, 90, 120 minutes Default: 30 minutes Adult Initial Cuff Inflation This parameter allows you to set the default NIBP cuff inflation pressure (the pressure to which the NIBP cuff inflates at the beginning of its first measurement after power up).
NIBP Settings (Optional) NIBP Patient Mode at Power Up This parameter allows you to set the default patient mode at power up. Values are as follows: Possible Values: Adult, Pediatric, Neonate Default: Adult NIBP Button Press & Hold This parameter allows you to define which NIBP function is activated when the front panel NIBP button ( ) is pressed and held for at least 2 seconds.
Chapter 2 Configuration Settings EtCO2 Settings (Optional) EtCO2 configuration parameters are available for update only if the EtCO2 monitoring option is installed on the unit and include: • Default EtCO2 Units • Displayed Zoom Level • Default EtCO2 Comp Setting • Power Save Mode at Power Up • EtCO2 Average On Power Up Default EtCO2 Units This parameter allows you to set the EtCO2 unit of measurement.
EtCO2 Settings (Optional) Power Save Mode at Power Up When this parameter is set to Yes, the unit automatically enters EtCO2 power save mode when powered on with a CAPNOSTAT sensor connected. The user must then enable the EtCO2 device using the Param/EtCO2 softkey menu before use. Values are as follows: Possible Values: Yes, No Default: Yes EtCO2 Average On Power Up This parameter sets the default time period over which the EtCO2 values are averaged.
Chapter 2 Configuration Settings Printer & Report Settings The Printer and Report Config settings are attributes that control what and how things are printed: • Set Report Restart Delays • Auto Generate Strips • Print 3 Leads When Leads Sel. Note: Printing is not available in AED mode. Set Report Restart Delays This setting allows you to review or change the length of time the device must be turned off before it automatically erases all summary report and full disclosure data.
Readiness Test Settings Readiness Test Settings The Readiness Test settings are attributes that control how and when the R Series unit is selftested. It includes the following settings: • Automatic Readiness Test • Readiness Test Time • Print Readiness Test Results • Enable Readiness Test Audible Alert (Future Use) • One Step Padz Installed Automatic Readiness Test This setting allows you to specify whether the R Series unit performs the automatic device check.
Chapter 2 Configuration Settings Enable Readiness Test Audible Alert (Future Use) This setting allows you to specify whether the unit emits a beep when the automatic device check fails. Values are as follows: Possible Values: No Default: No One Step Padz Installed This setting allows you to specify whether the R Series unit tests the R Series OneStep electrodes as part of the readiness test check, and which electrodes it tests.
AED Mode AED Mode The AED parameters are attributes that affect functions related to performing defibrillation with an R Series unit in AED mode and include: • Display ECG in AED Mode • Display HR in AED Mode • Manual Confirm Enable • Extra CPR Interval Before 1st Analysis • Duration of Extra CPR Interval • Display Remaining CPR Timer Display ECG in AED Mode This parameter allows you to set the unit to either display or not display the patient’s ECG trace on the monitor while in AED mode.
Chapter 2 Configuration Settings Extra CPR Interval Before 1st Analysis This parameter pertains only to AED mode. When this parameter is set to No (default), and the unit detects a pad connection upon poweron, it automatically starts an ECG analysis. If the unit is powered on with no pads connected, the unit displays the ATTACH PADS prompt until pads are connected. Once pads are connected, the unit automatically starts an ECG analysis.
Wi-Fi Settings (Optional) Wi-Fi Settings (Optional) The Wi-Fi settings are attributes that control how files are transferred and the security method that is used.
Chapter 2 Configuration Settings Local IP Address This setting allows you to specify the defibrillator’s IP address used in the static configuration. Note: This setting is available only when DHCP/Static IP is set to Static IP. Values are as follows: Possible Values: XXX.XXX.XXX.XXX where XXX is 000 to 255 Default: 192.168.100.100 Procedure Follow these steps to set up a Local IP Address. Step Action 1. From the Select Configuration Categories screen, select Wi-Fi, then press the Enter softkey.
Wi-Fi Settings (Optional) Gateway IP Address This setting allows you to specify the router’s IP used in the static configuration. Note: This setting is accessible only when DHCP/Static IP is set to Static IP. Values are as follows: Possible Values: XXX.XXX.XXX.XXX where XXX is 000 to 255 Default: 192.168.100.001 Procedure Follow these steps to set up a Gateway IP Address. Step Action 1. From the Select Configuration Categories screen, select Wi-Fi, then press the Enter softkey.
Chapter 2 Configuration Settings Subnet Mask This setting allows you to specify the subnet mask used in the static configuration. Note: This setting is accessible only when DHCP/Static IP is set to Static IP. Values are as follows: Possible Values: XXX.XXX.XXX.XXX where XXX is 000 to 255 Default: 255.255.255.000 Procedure Follow these steps to set up a subnet mask. Step Action 1. From the Select Configuration Categories screen, select Wi-Fi, then press the Enter softkey.
Wi-Fi Settings (Optional) Use DNS This setting allows you to specify whether to use a DNS to obtain the server IP address from the server name. Values are as follows: Possible Values: Yes, No Default: Yes Primary/Secondary DNS IP Address This setting allows you to specify the Primary and Secondary IP address of the DNS server to use if the following conditions are met: • Use DNS setting is set to Yes • DHCP/Static IP setting is set to Static IP Values are as follows: Possible Values: XXX.XXX.XXX.
Chapter 2 Configuration Settings Step 3. Action Use the softkey as follows: • — moves the highlight left and right through the character selection table. • Enter Char — enters the highlighted character for each digit of the IP address. • Return — returns you to the Wi-Fi Configuration screen. Press the Return softkey when you are finished entering the address. Follow the same steps to enter a Secondary DNS IP address.
Wi-Fi Settings (Optional) Step Action 3. In the character selection table: • moves the cursor to the next line. • . . . — changes the table to a different set of characters, including uppercase letters. Use the softkeys as follows: — moves the highlight up, down, left, and right through the character selection table. • Enter Char — enters the highlighted character for each digit of the SSIDs. • Return — returns you to the Wi-Fi Configuration screen.
Chapter 2 Configuration Settings Action 2. Press the Next Item softkey until Server Name (Full Disclosure) appears and press the Change Value softkey. The Set Server Name (Full Disclosure) screen appears. System Configuration Set Server Name (Full Disclosure) ZOLLDNS b c d e f g h i j k n o p q r s t u v w z 0 1 2 3 4 5 6 7 8 , @ . + - _ ? ‘ “ : l x 9 Enter Char 3. Cancel Del ... < a m y ; < < Step < Return In the character selection table: • moves the cursor to the next line. • . . .
Wi-Fi Settings (Optional) Procedure Follow these steps to set up a server IP address. Step Action 1. From the Select Configuration Categories screen, select Wi-Fi, then press the Enter softkey. System Configuration The Wi-Fi Settings screen appears. Wi-Fi Configuration Full Disclosure Transfer Mode Server Defib History Transfer Mode Both DHCP/ Static IP Static IP Page 1 of 7 Change Value 2.
Chapter 2 Configuration Settings Procedure Follow these steps to set up a server port name. Step Action 1. From the Select Configuration Categories screen, select Wi-Fi, then press the Enter softkey. System Configuration The Wi-Fi Settings screen appears. Wi-Fi Configuration Full Disclosure Transfer Mode Server Defib History Transfer Mode Both DHCP/ Static IP Static IP Page 1 of 7 Change Value 2.
Wi-Fi Settings (Optional) Procedure Follow these steps to set up a server name. Step Action 1. From the Select Configuration Categories screen, select Wi-Fi, then press the Enter softkey. System Configuration The Wi-Fi Settings screen appears. Wi-Fi Configuration Full Disclosure Transfer Mode Server Defib History Transfer Mode Both DHCP/ Static IP Static IP Page 1 of 7 2. Press the Next Item softkey until Server Name (Defib History) appears and press the Change Value softkey.
Chapter 2 Configuration Settings Values are as follows: Possible Values: XXX.XXX.XXX.XXX where XXX is 0 to 255 Default: 000.000.000.000 Procedure Follow these steps to set up an server IP address. Step Action 1. From the Select Configuration Categories screen, select Wi-Fi, then press the Enter softkey. System Configuration The Wi-Fi Settings screen appears.
Wi-Fi Settings (Optional) Server Port (Defib History) This setting allows you to specify the TCP port used to connect with the server if Defib History Transfer Mode is not set to None. Values are as follows: Default: 11000 Procedure Follow these steps to set up a server port. Step Action 1. From the Select Configuration Categories screen, select Wi-Fi, then press the Enter softkey. System Configuration The Wi-Fi Settings screen appears.
Chapter 2 Configuration Settings Passphrase/Keys (up to 2 entries) This setting allows you to specify the Wi-Fi passphrase. The ... key allows you to select upper case letters and other symbols. You can specify up to 2 entries with a 63-character limit. Procedure Follow these steps to set up a passphrase. Step Action 1. From the Select Configuration Categories screen, select Wi-Fi, then press the Enter softkey. System Configuration The Wi-Fi Settings screen appears.
Wi-Fi Transmission Mode Options Wi-Fi Transmission Mode Options Enable Report ID This setting allows you to enable or disable the option to enter a unique alphanumeric patient ID before Wi-Fi transmission of a Full Disclosure report. If set to No, the report will be sent automatically. Values are as follows: Possible Values: Yes, No Default: No Wi-Fi Region This setting allows you to specify the Wi-Fi domain (United States, Japan, or Europe).
Chapter 2 Configuration Settings 68 9650-0902-01 Rev.