ECG Settings
9650-0902-01 Rev. K 19
ECG Settings
ECG monitoring settings are attributes associated with using the R Series unit for ECG
monitoring, and include:
Notch Filter
Display Lead on Power-Up
Frequency Response
Gain on Power Up
Enable Leads Off
Enable Pacer Detection
Notch Filter
This setting allows yo u to select the ECG Notch Filter frequency for proper ac mains
interference rejection.
Values are as follows:
Display Lead on Power-Up
This setting allows you to select the lead to be displayed when you power on the unit.
Values are as follows:
Frequency Response
This setting allows yo u to select the ECG filter bandwidth to be used during ECG monitoring.
You can se t the unit’s frequen cy to either help reduce noise, he lp with the diagn ostic response,
or set it as a default response.
Values are as follows:
Possible Values: 50 Hz, 60 Hz
Default: 60 Hz
Possible Values: Paddles/Pads, Lead I, Lead II, Lead III
Default: Paddles/Pads
Possible Values: 0.5 - 21 Hz, 0.5 - 40 Hz, 1.0 - 21 Hz
Default: 0.5 - 21 Hz