Operator`s manual
[1] STAT PACE [2] DIVERT THERAPY [3] STAT SHOCK [4] PROGRAM [5] INTERROGATE
Figure 8. Right-side key pad
The following description o
f the right-side keypad corresponds to the labels
in the illustration (Figure 8 on page 21). The PRM must be in telemetry
communication with the pulse generator for these functions to be available.
• [1] Press STAT PACE to initiate emergency bradycardia paci ng at
predetermined high-o utput parameters.
• [2] Press DIVERT THERAPY to divert tach ycardi a therapy delivery.
• [3] Press STAT SHOCK to ini
tiate the delivery of an emergency maximum
energy shock.
• [4] Press PROGRAM to trans
mit new parameter values to the pulse
generator.
• [5] Press INTERROGATE to
obtain information stored in the pulse
generator memory.
[1] Speed keys [2] Paper-feed key [3] Calibration key [4] Baseline Key
Figure 9. Left-side keypad
The following descriptio n of the left-side keyp ad corresponds to the labels in
the illustration (Figure 9 on page 21).
• [1] Press the spee d keys to specify the paper speed for the internal
printer/recorder. The printout will show the date and time, lead(s)
being printed, gain setting, chart speed, and filter setting. To stop the
printer/recorder, press the speed key labeled “0” (zero).
21