Operator`s manual

Table 7. Guidance and manufacturers declaration - electromagnetic immunity -
for equipment and systems that are no t life supporting (continued)
Immunity test
IEC 60601 test
level
Compliance le vel
Electromagnetic
environment
a
guidance
b
Field strengths
from xed RF
transmitters, as
determined by an
electromagnetic
site survey
d
should
be less than the
compliance level
in each frequency
range.
e
Interference may
occur in the vicinity
of equipment
marked with the
following symbol:
a. The PRM and ZWT are intended for use in the electromagnetic environment specied in the table.
The customer or the user sh
ould ensure they are used in such an environment.
b. These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reection from structures, objects, and people.
c. At 80 MHz and 800 MHz, the hig
her frequency range applies.
d. Field strengths from xed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast
cannot be predicted theor
etically with accuracy. To assess the electromagnetic environment due
to xed RF transmitters, an electromagnetic site survey should be considered. If the measured
eld strength in the location in which the PRM and ZWT are used exceeds the applicable RF
compliance level shown i
n the table, they should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting
or relocating the PRM and ZWT.
e. Over the frequency range
150kHzto80MHz,eld strengths should be less t han 3 V/m.
Table 8. Recommended separation distances between portable and mobile RF
communications equipme
nt, and the PRM / ZWT
Separation distance acc
ording to frequency o f transmitter
a
m
Rated maximum
output power of
transmitter
bc
W
150 kHz to 80 MHz
d=1.2P
80 MHz to 800
MHz
d
d=1.2P
800 MH z to 2.5
GHz
d=2.3P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
39