6.2.5 Display
The display interface allows setting several display parameters.
These parameters are:
Spectrum: 1/1 Octave, 1/3 Octave or FFT;
Weight: A, C or Z (no weight);
Bandwidth: Low and high frequency range to consider. The global level is computed on this
frequency range;
Displayed Value:
Pressure, Intensity or PI index;
Global level or single frequency;
Color Scale:
Manual Range: the user sets the maximum and minimum values of the color range;
All Data: the maximum and minimum values in the map are used to define the
maximum and minimum values of the color range;
10, 5 and 3 dB MS: Sets the color range to the most significant dB of the map. As an
example, if the maximum value of the map is 93.8 dB and the 5 dB (MS) setting is
selected, the color scale will range from 88.8 to 93.8 dB.
Display Sign: Allows the user to select the displayed intensity sign :
Positive Only;
Negative Only;
Both Signs (Absolute).
Transparency: Sets the transparency of the sound map over the background image;
View Grid: Displays the computing grid on the map;
View Points: Displays the measurement points on the map;
View Tracks: Displays the scanning tracks on the map;
View Triangles: Displays the triangular interpolation structure on the map.