Technical data

22
86100A/B/C
Firmware Release Notes
changing modes
Measurement limits do not work when restored from saved setups
Mask test align failure does not cause an error in the queue
Mask margins not working with CG-GS waveforms
Markers could be moved outside of the window
Incorrect max string length reported with remote SYST:DSP command
Inconsistent behavior when saving a file to a full floppy
Horizontal Scale/Division setting didn't do anything
Can't remove unused measurement from measurement results tab
Binblock download of versus function does not work
Autoscale resets transducer conversion factors
Autoscale hung when acquisition stopped before using autoscale
Autoscale failure on math waveform turns on other waveforms
:VIEW CGM does not work after BLANk if a measurement command is sent first
:DISplay:DATa? returns inconsistent color scheme
:DISK:LOAD returns error 46 if in Oscilloscope mode
TDR fails vertical calibration if Threshold is set to "Units"
RZ eye width algorithm seems to place L open too close to mean; modified algorithm for
unsymmetrical edges
Module Delta Temperature reflects Mainframe temperature, not module
Module calibration time comes from 2 different sources; problem during daylight savings
time change
Measurement Limit Test still displays measurement after it's been turned off
Histogram tab stops displaying histogram data after math function
Delta Y now only shows value in Measurement Results tab when the units are the same
Databases dump when switching between RZ & NRZ measurements
Cannot select CGM as source with :WAVEFORM:SOURCE+B34 command
Cannot change channel offset for differential TDR channels
:WAVEFORM:DATA? Returns incorrect values in ASCII mode
:MEAS:CGRADE:JITTER? Does not return correct values; wasn't using source specified in
:MEAS:SOURCE
:CHAN#:UNIT:ATT, :CHAN#:UNIT:OFF and :CHAN#:PROBE now blocked when stimulus
is on
A.02.20 Release Notes, August 2, 2001 (Differences from Release
A.02.10)
New Feature List
Japanese Language Help
Support for 86106B
Support for 86109B
Support for 86102U
Defects Fixed:
Fixed defect that prevents the mask margins from being enabled when a setup file
containing margin information is loaded.
Fixed defect that forced users to reload masks after performing an autoscale while running
mask test.
The acquisition speed while eye measurements are turned on, which had been previously
degraded by a factor of four or five, is now fixed.
GPIB disk store goes to setup directory instead of specified directory, this is now fixed.
Resolved a problem with the RZ remote query for contrast ratio which causes the