Instruction Manual

MicroShockDetector plus
- 41 -
6.2 C A R R Y I N G O U T U P D A T E
After you have started your MONILOG
®
device in the update mode, start the PC software. The PC
software should automatically detect the MONILOG
®
device which is in the update mode and open
a window to select an „Encrypted Hex File“.
After selecting the update file the update process, which must not be interrupted, starts. The red
STATUS LED on the device lights up continuously.
Figure 31: Firmware update
If the progress bar shows 100%, then the message of the successful completion of the update
appears.
Figure 32: Completion of the update process
The USB connection must now be disconnected. The red STATUS LED on the device goes out. The
device must be switched on again.
If the update process fails, a corresponding error message is displayed. The MONILOG
®
device
switches off independently. Due to the update interruption the device firmware could not be
completely installed. In some circumstances, the device is not switchable any more in the normal
operating mode due to this. Please repeat the update process according to the instructions of the
entire chapter 6 Firmware update“.