3. To change the measurement direction for the measurement point, press F2 XYZ.
Each time you press F2 XYZ, the selected direction changes.
8.7.3 Change the polarity
The polarity indicates the direction to collect the measurement. The measurement
direction is usually determined by the construction of the equipment. If you cannot orient
the sensor in the positive direction, change the polarity to a negative direction
measurement.
You can change the polarity after you collect data for the measurement point.
Procedure
1.
Activate a job.
2. From the ODS/Modal main menu, use the arrows or F9 Point Search key to go to the
desired measurement direction.
3. Press F3 Change Sign +/-.
The Point field changes to display the updated measurement direction.
8.7.4 Delete the data collected for a measurement point
Note
Use caution with the Clear Data option. Deleted data cannot be recovered.
Procedure
1.
Activate a job.
2. From the ODS/Modal main menu, use the up and down arrow keys or the F9 Point
Search key to select a measurement point.
3. Press F5 Clear Data.
4. Press Enter.
The analyzer clears the data from only the active point.
8.7.5 Redo a measurement
You can store one set of data per measurement point/direction. However, you can
overwrite a measurement.
Procedure
1. From the ODS/Modal main menu, press F9 Point Search or use the arrows on the
keypad to go to a measurement point.
2. Press Enter to collect data.