User's Manual

Table Of Contents
Procedure
1.
Copy the new programs to a folder on your computer.
2. Use the appropriate cable to connect your analyzer to a computer where AMS
Machinery Manager is installed.
3. On the computer, open Data Transfer.
4. In the Options pane, set the Key Table and Firmware fields to the firmware folder
you created in step 1.
a. Click the Options icon.
b. Click Browse next to the Key Table field and select the folder containing the
firmware.
c. Click Browse next to the Firmware field and select the folder containing the
firmware.
d. Click Apply.
5. On the analyzer, press Home > F6 Program Manager > F8 Connect For Transfer.
The analyzer displays a message indicating "Host Computer Connected" and lists the
options under the Programs Available for Download section.
Note
If the following message appears, you must update your firmware before continuing: "The
host computer has a newer version of the base firmware. You should follow the procedure in
your user's manual for loading new base firmware before you load any new programs". The
analyzer will not open any programs until you update the firmware. The analyzer may prompt
you to update the firmware after you load or update the programs.
6. Press F11 Program or F12 Program or use the up and down arrow keys to highlight a
program, and press F10 Select/Unselect Program to select a program.
7.
Press F9 Start Download.
Wait until all programs in the analyzer are up to date. The download typically takes a
few minutes.
8. Press Enter or Reset.
Postrequisites
View the version number for the firmware and programs to ensure the updates were
successful.
Transfer files with CSI 2140
49
DRAFT